Function and Applications
Self recovering Halogenid countertube for detection of Alpha-,Beta- und Gamma-radiation.
Benefits
mounted in metal cylinder with fixed 500 mm long BNC-cable
Including protection cap for countertube
Equipment and technical data
Mica window
Density of Mica window:2...3 mg/cm²
Operation voltage: 500 V
Plateau length: 200 V
Plateau slope: 0,04%/V
Death time: approx. 100 µs
Zero rate: approx. 15 Pulse/min
Diameter of housing: 22 mm
Diameter of counter tube: 15 mm
Counter tube length: 76 mm
Mass: 103 g
Self recovering Halogenid countertube for detection of Alpha-,Beta- und Gamma-radiation.
Benefits
mounted in metal cylinder with fixed 500 mm long BNC-cable
Including protection cap for countertube
Equipment and technical data
Mica window
Density of Mica window:2...3 mg/cm²
Operation voltage: 500 V
Plateau length: 200 V
Plateau slope: 0,04%/V
Death time: approx. 100 µs
Zero rate: approx. 15 Pulse/min
Diameter of housing: 22 mm
Diameter of counter tube: 15 mm
Counter tube length: 76 mm
Mass: 103 g
Properties
- P0900500